L'unità PTB330S combina un barometro PTB330 con un indicatore portatile MI70 (che alimenta gli strumenti e fornisce il display per visualizzare i dati) ed (opzionale) una sonda HMP155 per temperatura ed umidità. Lo strumento è adatto ad effettuare misure di riferimento in ambito meteorologico e industriale; viene dotato di valigetta da trasporto weatherproof per una facile movimentazione.

 

Specifiche Tecniche

Barometric Pressure range 500...1100 hPa
Barometric Pressure linearity ±0.05 hPa
Barometric Pressure hysteresis ±0.03 hPa
Barometric Pressure repeatability ±0.03 hPa
Barometric Pressure calibration uncertainty ±0.07 hPa
Barometric Pressure accuracty at +20 °C ±0.10 hPa
Barometric Pressure temperature dependence ±0.01 hPa
Barometric Pressure total accuracy at -40...+60 °C ±0.15 hPa
Barometric Pressure long-term stability (/year) ±0.1 hPa
Barometric Pressure response time 2 s
Humidity measurement range 0...100% RH
Humidity accuracy at +15...+25 °C, 0...90% RH ±1% RH
Humidity accuracy at +15...+25 °C, 90...100% RH ±1.7% RH
Humidity accuracy at -10...+40 °C ±(1.0 + 0.008 x reading)% RH
Humidity Factory Calibration Uncertainty (+20 °C), 0...40% RH ±0.6% RH
Humidity Factory Calibration Uncertainty (+20 °C), 40...97% RH ±1.0% RH
Humidity response time (63%) at +20 °C in still air, with a sintered PTFE filter 20 s
Humidity response time (90%) at +20 °C in still air, with a sintered PTFE filter 60 s
Temperature measurement rage -10...+40 °C
Temperature accuracy at -10...+20 °C ±(0.176 - 0.0028 x temperature) °C
Temperature accuracy at +20...+40 °C ±(0.07 + 0.0025 x temperature ) °C
Temperature response time (63%) with additional temperature probe in 3 m/s air flow < 20 s
Temperature response time (90%) with additional temperature probe in 3 m/s air flow < 35 s
Power supply Rechargeable NiMH battery pack with AC-adapter or 4xAA-size alkalines, type IEC LR6
MI70 probe ports 2
MI70 data interface RS-232
PTB330 supply voltage 10...35 VDC (if not powered by MI70)
PTB330 data interface RS-232C
HMP155 data interface RS-485
Operating temperature range -10...+40 °C

 

Per maggiori dettagli sul prodotto, download aggiuntivi, FAQ, accessori ed esempi di applicazioni visitare il sito del produttore alla pagina PTB330S

Brochure PTB330S Data Sheet PTB330S Manuale PTB330S